0315 Metallic Powders Gustavsson Flow Meter is produced according to ISO 13517. It is used to measure the flow rate of the metallic powder and mixes.
Principle
Measurement of the time required for 50 g of a metallic powder to flow through the orifice of a calibrated funnel of standardized dimensions.
Apparatus
-Calibrated funnel
-Stand and horizontal vibration-free base
-Balance
-Stopwatch
-Chinese emery grit
(Apparatus is subject to quotation)
Calibration of the funnel
-Calibration by the manufacturer of the funnel(AimSizer)
-Calibration by the user of the funnel(Validation)